Abstract
Improvement of a conventional Jeol Raman spectrometer with a single channel photo multiplier detector is described. New optical components (fibres, mirror, lens and CCD detector) have been chosen to design a high quality and easy-to-use instrument. Tests have shown that with this modified spectrometer Raman spectra can be acquired of a quality comparable to the spectra obtained previously, but the time needed to obtain a spectrum is markedly reduced. Selected test spectra and a simple calibration procedure to obtain the wavenumber values from the band CCD pixel position are presented.
Original language | English |
---|---|
Journal | Applied Spectroscopy Reviews |
Volume | 39 |
Issue number | 3 |
Pages (from-to) | 385-397 |
ISSN | 0570-4928 |
Publication status | Published - 2004 |