Technical validation of the Di3D stereophotogrammetry surface imaging system

R.J. Winder, Tron Andre Darvann, W. McKnight, J.D.M. Magee, P. Ramsay-Baggs

    Research output: Contribution to journalJournal articleResearchpeer-review

    Abstract

    The purpose of this work was to assess the technical performance of a three-dimensional surface imaging system for geometric accuracy and maximum field of view. The system was designed for stereophotogrammetry capture of digital images from three-dimensional surfaces of the head, face, and neck. A mannequin head was prepared for imaging by adding texture in the form of red paint, and facial landmarks as black ink dots. The mannequin was imaged at the manufacturer's recommended settings for human studies. Colour-coded surface difference images among repeated exposures were computed. We compared measurements of physical linear distance with digital measurements. The three-dimensional stereophotogrammetry system had a mean error in the three-dimensional surfaces of 0.057 mm, a repeatability error (variance) of 0.0016 mm, a mean error of 0.6 mm in linear measurements compared with manual measurements, and a field of view of 170 degrees horizontally and 102 degrees vertically.
    Original languageEnglish
    JournalBritish Journal of Oral and Maxillofacial Surgery
    Volume46
    Issue number1
    Pages (from-to)33-37
    ISSN0266-4356
    DOIs
    Publication statusPublished - 2008

    Keywords

    • three-dimensional surface
    • stereophotogrammetry
    • geometric validation
    • facial morphometry

    Fingerprint Dive into the research topics of 'Technical validation of the Di3D stereophotogrammetry surface imaging system'. Together they form a unique fingerprint.

    Cite this