Skip to main navigation Skip to search Skip to main content

Supplementary comparison CCRI(I)-S3 of standards for absorbed dose to water in 60Co gamma radiation at radiation processing dose levels

  • M. R. McEwen
  • , P. H.G. Sharpe
  • , I. M. Pazos
  • , A. Miller
  • , E. Pawlak
  • , S. Ninlaphruk
  • , Y. Zhang
  • , C. Kessler
  • National Research Council of Canada
  • National Physical Laboratory
  • National Institute of Standards and Technology
  • Comisión Nacional de Energía Atómica
  • National Institute of Metrology China
  • Bureau international des poids et mesures

Research output: Contribution to journalJournal articleResearchpeer-review

Abstract

A supplementary comparison of standards for absorbed dose to water in 60Co fields used for calibrations at radiation processing dose levels has been completed. Alanine dosimeters from the NIST and the NPL were used as transfer dosimeters and all participant irradiations were carried out in self-shielded irradiators. Irradiations were also carried out at the BIPM to allow direct comparison of dose ratios with the Key Comparison BIPM.RI(I)-K4. No significant difference was seen between the dose ratios obtained using the NPL and NIST alanine systems, no significant impact of mailing on dosimeter response was noted, and the adopted protocol limited any dose rate effects to the level of Type A uncertainties. The national standards of the participants are in agreement within the standard uncertainties, which are in the range from 1 to 2 parts in 102.

Original languageEnglish
Article number06012
JournalMetrologia
Volume59
Issue number1A
Number of pages12
ISSN0026-1394
DOIs
Publication statusPublished - 2022

Fingerprint

Dive into the research topics of 'Supplementary comparison CCRI(I)-S3 of standards for absorbed dose to water in 60Co gamma radiation at radiation processing dose levels'. Together they form a unique fingerprint.

Cite this