Abstract
A supplementary comparison of standards for absorbed dose to water in 60Co fields used for calibrations at radiation processing dose levels has been completed. Alanine dosimeters from the NIST and the NPL were used as transfer dosimeters and all participant irradiations were carried out in self-shielded irradiators. Irradiations were also carried out at the BIPM to allow direct comparison of dose ratios with the Key Comparison BIPM.RI(I)-K4. No significant difference was seen between the dose ratios obtained using the NPL and NIST alanine systems, no significant impact of mailing on dosimeter response was noted, and the adopted protocol limited any dose rate effects to the level of Type A uncertainties. The national standards of the participants are in agreement within the standard uncertainties, which are in the range from 1 to 2 parts in 102.
| Original language | English |
|---|---|
| Article number | 06012 |
| Journal | Metrologia |
| Volume | 59 |
| Issue number | 1A |
| Number of pages | 12 |
| ISSN | 0026-1394 |
| DOIs | |
| Publication status | Published - 2022 |
Fingerprint
Dive into the research topics of 'Supplementary comparison CCRI(I)-S3 of standards for absorbed dose to water in 60Co gamma radiation at radiation processing dose levels'. Together they form a unique fingerprint.Cite this