Stationary echo canceling in velocity estimation by time-domain cross-correlation

    Research output: Contribution to journalJournal articleResearchpeer-review

    643 Downloads (Pure)

    Abstract

    The application of stationary echo canceling to ultrasonic estimation of blood velocities using time-domain cross-correlation is investigated. Expressions are derived that show the influence from the echo canceler on the signals that enter the cross-correlation estimator. It is demonstrated that the filtration results in a velocity-dependent degradation of the signal-to-noise ratio. An analytic expression is given for the degradation for a realistic pulse. The probability of correct detection at low signal-to-noise ratios is influenced by signal-to-noise ratio, transducer bandwidth, center frequency, number of samples in the range gate, and number of A-lines employed in the estimation. Quantitative results calculated by a simple simulation program are given for the variation in probability from these parameters. An index reflecting the reliability of the estimate at hand can be calculated from the actual cross-correlation estimate by a simple formula and used in rejecting poor estimates or in displaying the reliability of the velocity estimated
    Original languageEnglish
    JournalI E E E Transactions on Medical Imaging
    Volume12
    Issue number3
    Pages (from-to)471-477
    ISSN0278-0062
    DOIs
    Publication statusPublished - 1993

    Bibliographical note

    Copyright: 1993 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEE

    Fingerprint

    Dive into the research topics of 'Stationary echo canceling in velocity estimation by time-domain cross-correlation'. Together they form a unique fingerprint.

    Cite this