Skip to main navigation Skip to search Skip to main content

Scalar Calibration of Vector Magnetometers

    Research output: Contribution to journal › Journal article › Research › peer-review

    Abstract

    The calibration parameters of a vector magnetometer are estimated only by the use of a scalar reference magnetometer. The method presented in this paper differs from those previously reported in its linearized parametrization. This allows the determination of three offsets or signals in the absence of a magnetic field, three scale factors for normalization of the axes and three non-orthogonality angles which build up an orthogonal system intrinsically in the sensor. The advantage of this method compared with others lies in its linear least squares estimator, which finds independently and uniquely the parameters for a given data set. Therefore, a magnetometer may be characterized inexpensively in the Earth's magnetic-field environment. This procedure has been used successfully in the pre-flight calibration of the state-of-the-art magnetometers on board the magnetic mapping satellites Orsted, Astrid-2, CHAMP and SAC-C. By using this method, full-Earth-field-range magnetometers (+/-65536.0 nT) can be characterized down to an absolute precision of 0.5 nT, non-orthogonality of only 2 arcsec and a resolution of 0.2 nT.
    Original languageEnglish
    JournalMeasurement Science and Technology
    Volume11
    Issue number2
    Pages (from-to)120-132
    ISSN0957-0233
    Publication statusPublished - 2000

    Keywords

    • linearization
    • tri-axial magnetometer
    • least squares
    • calibration

    Fingerprint

    Dive into the research topics of 'Scalar Calibration of Vector Magnetometers'. Together they form a unique fingerprint.

    Cite this