Radiative MRI Coil Design Using Parasitic Scatterers: MRI Yagi

Juan D. Sanchez-Heredia, Johan Avendal, Adnan Bibic, Buon Kiong Lau*

*Corresponding author for this work

Research output: Contribution to journalJournal articleResearchpeer-review

412 Downloads (Pure)

Abstract

Conventionally, radiofrequency (RF) coils used for magnetic resonance imaging (MRI) are electrically small and designed for nearfield operation. Therefore, existing antenna design techniques are mostly irrelevant for RF coils. However, the use of higher frequencies in ultrahigh field (UHF) MRI allows for antenna design techniques to be adapted to RF coil designs. This study proposes the use of parasitic scatterers to improve the performance of an existing 7T MRI coil called the single-sided adapted dipole (SSAD) antenna. The results reveal that scatterers arranged in a Yagi fashion can be applied to reduce local specific absorption rate (SAR) maxima of a reference SSAD by 40% with only a 6% decrease in the propagated B1 + field at the tissue depth of 15 cm. The higher directivity of the proposed design also decreasing the coupling with additional elements, making this antenna suitable for use in high density arrays. These findings show the potential of parasitic scatterers as an effective method to improve the performance of existing radiative MRI coils.

Original languageEnglish
JournalIEEE Transactions on Antennas and Propagation
Volume66
Issue number3
Pages (from-to)1570 - 1575
ISSN0018-926X
DOIs
Publication statusPublished - 19 Jan 2018

Keywords

  • Antenna arrays
  • Dipole antennas
  • Electromagnetic propagation in absorbing media
  • Magnetic resonance imaging
  • Radio frequency
  • Specific absorption rate
  • Substrates
  • Yagi-Uda antennas

Fingerprint Dive into the research topics of 'Radiative MRI Coil Design Using Parasitic Scatterers: MRI Yagi'. Together they form a unique fingerprint.