Quantification of MRI visibility and artifacts at 3T of liquid fiducial marker in a pancreas tissue-mimicking phantom

Sergej Schneider*, Rasmus Irming Jølck, Esther Gera Cornelia Troost, Aswin Louis Hoffmann

*Corresponding author for this work

    Research output: Contribution to journalJournal articleResearchpeer-review

    Abstract

    Purpose: X-ray-based position verification of the target volume in image-guided radiation therapy (IGRT) of patients with pancreatic ductal adenocarcinoma (PDAC) is currently performed on solid fiducial markers that are implanted under endoscopic ultrasonography. A new biodegradable liquid fiducial marker has recently been introduced. To assess its potential use for magnetic resonance imaging (MRI)-guided photon or proton radiotherapy of PDAC, the MRI visibility and artifacts of this marker were quantified and compared against solid gold markers.Material and methods: Different spherical volumes (10 mu L, 25 mu L, 50 mu L, and 100 mu L) of a biodegradable liquid fiducial marker as well as seven differently sized and oriented solid gold (0.35 mm diameter; 5 mm and 10 mm length) and iron-gold alloy fiducial markers (0.28 mm diameter; 1 cm and 2 cm length) were implanted in a spherical gel phantom, mimicking the proton spin relaxation properties of healthy pancreatic tissue at 3 Tesla. MR relaxometry was performed to quantify the size and magnitude of the decrease in the effective transversal relaxation time T*(2) and relative proton density rho(H) as a measure of potential visibility and to quantify the size and magnitude of the increase in magnetic field inhomogeneity Delta B-0 as a measure of potential signal artifacts. The phantom was scanned in a 3.0 T PET/MR scanner with an eight-channel head coil.Results: The solid fiducial markers showed a direct linear relationship between the potentially visible size and artifact size. The liquid fiducial marker showed a tendency toward a potentially visible size at smaller artifacts. Liquid markers from 25 to 100 mu L generated visible volumes comparable to the size of the solid markers. The magnitude of visibility was the highest for the liquid fiducial marker with volumes of 25-100 mu L showing no correlation with the magnitude of artifact. The solid markers showed a strong nonlinear correlation between magnitude of visibility and artifact, whereas the solid marker consisting of a gold-iron alloy induced the strongest artifacts.Conclusion: The liquid fiducial marker causes signal voids on MRI due to its absence of water hydrogen atoms without strongly affecting the magnetic field in the surrounding tissue. The alteration of the static magnetic field was found to be the main effect leading to the visibility of the solid fiducial markers. Hence, especially when a low level of image distortion is required, MRI characteristics of the liquid marker surpass those of solid gold markers currently being used for IGRT of PDAC. (c) 2017 American Association of Physicists in Medicine
    Original languageEnglish
    JournalMedical Physics
    Volume45
    Issue number1
    Pages (from-to)37-47
    ISSN0094-2405
    DOIs
    Publication statusPublished - 2018

    Keywords

    • Artifacts
    • Fiducial marker
    • Image-guided radiation therapy
    • MRI
    • Visibility

    Fingerprint

    Dive into the research topics of 'Quantification of MRI visibility and artifacts at 3T of liquid fiducial marker in a pancreas tissue-mimicking phantom'. Together they form a unique fingerprint.

    Cite this