Parallel imaging of hyperpolarized 13C MRI using 23Na sensitivity profiles

Rie Beck Olin, Juan Diego Sanchez, James T. Grist, Wenjun Wang, Nikolaj Bøgh, Vitaliy Zhurbenko, Esben S. Hansen, Rolf F. Schulte, Damian John Tyler, Christoffer Laustsen, Jan Henrik Ardenkjær-Larsen

Research output: Contribution to conferenceConference abstract for conferenceResearchpeer-review

Abstract

Parallel imaging can aid hyperpolarized 13C MRI in extracting more information in the limited time window of the hyperpolarized signal. Using a dedicated, flexible 13C receive coil design with coupling coefficients matched for both 13C and 23Na, sensitivity profiles for reconstruction can be acquired at the 23Na frequency. We demonstrate this method in two hyperpolarized in vivo experiments involving pig kidneys and human brain using a two-times accelerated 3D blipped stack-of-spirals sequence with dual-resolution. The results show good SNR, coverage, and resolution. The method is promising for
integrating and automating parallel imaging for hyperpolarized 13C MRI.
Original languageEnglish
Publication date2022
Number of pages7
Publication statusPublished - 2022
Event2022 ISMRM & SMRT Annual Meeting & Exhibition - London, United Kingdom
Duration: 7 May 202212 May 2022

Conference

Conference2022 ISMRM & SMRT Annual Meeting & Exhibition
Country/TerritoryUnited Kingdom
CityLondon
Period07/05/202212/05/2022

Fingerprint

Dive into the research topics of 'Parallel imaging of hyperpolarized 13C MRI using 23Na sensitivity profiles'. Together they form a unique fingerprint.