Measurements of Charge Sharing Effects in Pixilated CZT/CdTe Detectors

    Research output: Chapter in Book/Report/Conference proceedingArticle in proceedingsResearchpeer-review

    Abstract

    In this paper, charge sharing and charge loss effects in pixilated CZT/CdTe detectors are investigated by measurements. We measured charge sharing effects function of the inter-pixel gap (with same pixel pitch), the photon energy and the detector bias voltage for a large numbers of CZT and CdTe pixel detector samples. The results are used for the development of the large area X-ray and Gamma ray detector for the Atmosphere-Space Interactions Monitor (ASIM) planned for the ISS ESA Columbus module. Charge sharing measurements on detector samples with identical size and pixel geometry with different inter-pixel gap showed charge loss on the gap surface which is linear dependent on gap size.
    Original languageEnglish
    Title of host publicationIEEE Nuclear Science Symposium
    PublisherIEEE
    Publication date2007
    Pages2252-2257
    ISBN (Print)978-1-4244-0922-8
    DOIs
    Publication statusPublished - 2007
    Event2007 IEEE Nuclear Science Symposium and Medical Imaging Conference - Honolulu, United States
    Duration: 28 Oct 20073 Nov 2007
    https://ewh.ieee.org/soc/nps/nss-mic/2007/

    Conference

    Conference2007 IEEE Nuclear Science Symposium and Medical Imaging Conference
    Country/TerritoryUnited States
    CityHonolulu
    Period28/10/200703/11/2007