In this paper, charge sharing and charge loss effects in pixilated CZT/CdTe detectors are investigated by measurements. We measured charge sharing effects function of the inter-pixel gap (with same pixel pitch), the photon energy and the detector bias voltage for a large numbers of CZT and CdTe pixel detector samples. The results are used for the development of the large area X-ray and Gamma ray detector for the Atmosphere-Space Interactions Monitor (ASIM) planned for the ISS ESA Columbus module. Charge sharing measurements on detector samples with identical size and pixel geometry with different inter-pixel gap showed charge loss on the gap surface which is linear dependent on gap size.
Original language | English |
---|---|
Title of host publication | IEEE Nuclear Science Symposium |
Publisher | IEEE |
Publication date | 2007 |
Pages | 2252-2257 |
ISBN (Print) | 978-1-4244-0922-8 |
DOIs | |
Publication status | Published - 2007 |
Event | 2007 IEEE Nuclear Science Symposium and Medical Imaging Conference - Honolulu, United States Duration: 28 Oct 2007 → 3 Nov 2007 https://ewh.ieee.org/soc/nps/nss-mic/2007/ |
Conference
Conference | 2007 IEEE Nuclear Science Symposium and Medical Imaging Conference |
---|---|
Country/Territory | United States |
City | Honolulu |
Period | 28/10/2007 → 03/11/2007 |
Internet address |