Abstract
In this work we demonstrate the construction of an ‘inside-out’ sensor geometry for electrical capacitance tomography (ECT). The inside-out geometry has the electrodes placed around a tube, as usual, but measuring ‘outwards’. The flow between the electrodes and an outer tube is reconstructed; allowing the inside-out sensor to move inside the outer tube. A test sensor was constructed and capacitances were measured using the charge transfer technique. Sensitivity matrices for the inside-out sensor were calculated with a finite element approach and some special issues with the sensitivity matrices are discussed. An adaptation of the Landweber algorithm, which works very well for the inside-out geometry, is presented and a definition of the spatial resolution of an ECT sensor is suggested. Tomograms from a test run of an inside-out sensor are presented and measurements of watercut are compared with images obtained by a camera and a simple direct result based on the capacitance vector.
Original language | English |
---|---|
Journal | Flow Measurement and Instrumentation |
Volume | 22 |
Issue number | 2 |
Pages (from-to) | 104-109 |
ISSN | 0955-5986 |
DOIs | |
Publication status | Published - 2011 |
Keywords
- Inside-out sensor geometry
- ETC
- Sensitivity matrix