Improved perfusion quantification in FAIR imaging by offset correction

  • Karam Sidaros
  • , Irene Klærke Andersen
  • , Henrik Gesmar
  • , Egill Rostrup
  • , Henrik B. W. Larsson

Research output: Contribution to journal › Journal article › Research › peer-review

Abstract

Perfusion quantification using pulsed arterial spin labeling has been shown to be sensitive to the RF pulse slice profiles. Therefore, in Flow-sensitive Alternating-Inversion Recovery (FAIR) imaging the slice selective (ss) inversion slab is usually three to four times thicker than the imaging slice. However, this reduces perfusion sensitivity due to the increased transit delay of the incoming blood with unperturbed spins. In the present article, the dependence of the magnetization on the RF pulse slice profiles is inspected both theoretically and experimentally. A perfusion quantification model is presented that allows the use of thinner ss inversion slabs by taking into account the offset of RF slice profiles between ss and nonselective inversion slabs. This model was tested in both phantom and human studies.
Original languageEnglish
JournalMagnetic Resonance in Medicine
Volume46
Issue number1
Pages (from-to)193-197
ISSN0740-3194
DOIs
Publication statusPublished - 2001

Fingerprint

Dive into the research topics of 'Improved perfusion quantification in FAIR imaging by offset correction'. Together they form a unique fingerprint.