Improved Field Homogeneity for Transmission Line MRI Coils Using Series Capacitors

Research output: Chapter in Book/Report/Conference proceedingArticle in proceedingsResearchpeer-review

403 Downloads (Pure)

Abstract

High field magnetic resonance imaging (MRI) systems often use short sections of transmission lines for generating and sensing alternating magnetic fields. Due to distributed nature of transmission lines, the generated field is inhomogeneous. This work investigates the application of series capacitors to improve the field homogeneity. The resulting magnetic field distribution is estimated analytically and evaluated numerically. The results are compared to a case of a conventional transmission line coil realization.
Original languageEnglish
Title of host publicationProceedings. 2015 SBMO/IEEE MTT-S International Microwave and Optoelectronics Conference (IMOC)
Number of pages5
PublisherIEEE
Publication date2015
ISBN (Print)978-1-5090-0431-7
DOIs
Publication statusPublished - 2015
Event2015 SBMO/IEEE MTT-S International Microwave and Optoelectronics Conference (IMOC) - Porto de Galinhas, Brazil
Duration: 3 Nov 20156 Nov 2015
http://www.imoc2015.com/

Conference

Conference2015 SBMO/IEEE MTT-S International Microwave and Optoelectronics Conference (IMOC)
CountryBrazil
CityPorto de Galinhas
Period03/11/201506/11/2015
Internet address

Keywords

  • Magnetic resonance imaging
  • MRI coil
  • Transmission line
  • TEM coil

Fingerprint

Dive into the research topics of 'Improved Field Homogeneity for Transmission Line MRI Coils Using Series Capacitors'. Together they form a unique fingerprint.