Geometrical Comparison Measures for Tumor Delineation, what do they mean for the Actual Dosis Plan?

Christian Hollensen, G. Persson, L. Højgaard, L. Specht, Rasmus Larsen

    Research output: Contribution to journalConference abstract in journalResearchpeer-review

    Abstract

    Purpose/Objective: Gross tumour volume (GTV) delineation is central for radiotherapy planning. It provides the basis of the clinical target volume and finally the planning target volume (PTV) which is used for dose optimization. GTV delineations are prone to intermethod and interobserver variation. In clinical studies this variation is commonly represented by geometrical volume comparison measures (GVCMs) as volume assessment, centre of mass and overlap. The correlation between these measures and the radiotherapy plan are however unclear. The aim of the present study is to investigate the correlation between GVCMs and the radiotherapy plans of patients with peripheral lung
    tumours.

    Materials and Methods: Peripheral lung tumours of 10 patients referred for stereotactic body radiotherapy in 2008 were delineated by 3 radiologists and 3 oncologists. From these GTV delineations 6 different radiotherapy plans with RapidArc© were created for each patient using the same procedure for creation of PTV and dose optimisation. For each patient the volume receiving 90 % of the prescribed dose (V90) and the minimum dose that 90 % of the volume receives (D90) was extracted for the 6 delineations on each of radiotherapy plans. GVCMs as Dice overlap coefficient, mismatch, volume difference, center of mass distance, and Haussdorff distance were extracted between each pair of the delineations of GTV for each patient. Mismatch was defined as the volume of a GTV delineation outside the GTV delineation used to create the PTV divided volume of the GTV used to create the PTV. The Pearson correlation between the GVCMs and their corresponding difference in V90 and D90 was calculated and their statistical difference from zero and each other was tested with a ttest with a pvalue of 0.05.

    Results: The V90 and D90 were extracted for the 6 different PTVs on the 60 radiotherapy plans. The standard deviation for V90 and D90 were 5.5 % of the volume and 4.1 Gy respectively, The standard deviation in one image plane of one patient can be seen in the figure.

    Figure: Standard deviation of the different radiotherapy plans in one plane for one patient. GTV contours in white. 150 estimations of the difference between the volumes were calculated for each of the GVCMs. The correlation results can be seen in the table.

    Table: Correlation coefficients
    Correlation of V90 Correlation of D90
    Dice Coefficient 0.44 ± 0.10 0.43 ± 0.10
    Mismatch 0.82 ± 0.03 0.71 ± 0.05
    Volume difference 0.37 ± 0.10 0.32 ± 0.10
    Center of mass distance 0.37 ± 0.09 0.49 ± 0.08
    Hausdorff distance 0.37 ± 0.10 0.37 ± 0.10

    All the correlation coefficients were found significantly different from 0. The correlation coefficient for mismatch was significantly different from all the other GVCMs for both V90 and D90. The correlation coefficient for center of mass was significantly different from the volume difference for D90.

    Conclusions: Mismatch between GTVs is significantly more correlated with V90 and D90 than other GVCMs. Mismatch between GTVs could be used as a indicator for difference in V90 and D90 of their corresponding radiotherapy plans.
    Original languageEnglish
    JournalRadiotherapy & Oncology
    Volume103
    Issue numberSupplement 1
    Pages (from-to)S308
    ISSN0167-8140
    DOIs
    Publication statusPublished - 2012
    EventESTRO 31 - Barcelona, Spain
    Duration: 9 May 201213 May 2012
    http://www.estro-events.org/pages/estro31.aspx

    Conference

    ConferenceESTRO 31
    Country/TerritorySpain
    CityBarcelona
    Period09/05/201213/05/2012
    Internet address

    Fingerprint

    Dive into the research topics of 'Geometrical Comparison Measures for Tumor Delineation, what do they mean for the Actual Dosis Plan?'. Together they form a unique fingerprint.

    Cite this