Free-field correction values for Interacoustics DD 45 supra-aural audiometric earphones

Torben Poulsen

    Research output: Contribution to journalJournal articleResearchpeer-review

    Abstract

    This paper report free-field correction values for the Interacoustics DD 45 audiometric earphones. The free-field correction values for earphones provide the loudness based equivalence to loudspeaker presentation. Correction values are especially used for the calibration of audiometric equipment for speech audiometry performed with headphones. Calibration values may be found in e.g. the ISO 389 series of standards. The free-field correction values were determined by means of loudness balance measurements of one-third octave noises (centre frequencies 125 Hz to 8000 Hz) presented alternately from a loudspeaker in a free field and from the earphones. The procedure was essentially in accordance with the free-field frequency response procedure described in IEC 60268-7: Headphones and earphones. The study sample consisted of four earphones and 14 test subjects. Free field correction values are reported for the acoustic coupler IEC 60318-3 (NBS 9-A) and for the ear simulator IEC 60318-1. The results are in good agreement with the results of another independent investigation. The reported free-field correction values may be used as part of the basis for future standardization of the DD 45 earphone.
    Original languageEnglish
    JournalInternational Journal of Audiology
    Volume50
    Issue number6
    Pages (from-to)361-366
    ISSN1499-2027
    DOIs
    Publication statusPublished - 2010

    Keywords

    • Audiometry
    • Hearing threshold
    • Calibration

    Fingerprint

    Dive into the research topics of 'Free-field correction values for Interacoustics DD 45 supra-aural audiometric earphones'. Together they form a unique fingerprint.

    Cite this