Abstract
We describe the design and performance of a 28-channel RF receive array optimized for human head 13C imaging at 3T. The array employs a fixed and symmetric cylindrical geometry to provide predictable prior information of the coil sensitivity profiles, which can be used to resolve practical challenges of hyperpolarized parallel imaging and to improve performance. The results show good SNR and spatial encoding of the different elements, with only small performance difference between the individual elements.
Original language | English |
---|---|
Title of host publication | Proceedings of the International Society for Magnetic Resonance in Medicine |
Number of pages | 4 |
Volume | 28 |
Publisher | The International Society for Magnetic Resonance in Medicine |
Publication date | 2020 |
Article number | 4106 |
Publication status | Published - 2020 |
Event | 2020 ISMRM & SMRT Virtual Conference & Exhibition - Online Duration: 8 Aug 2020 → 14 Aug 2020 |
Conference
Conference | 2020 ISMRM & SMRT Virtual Conference & Exhibition |
---|---|
Location | Online |
Period | 08/08/2020 → 14/08/2020 |