Abstract
A technique for estimating the full flow velocity vector has previously been presented by our group. Unlike conventional estimators, that only detect the axial component of the flow, this new method is capable of estimating the transverse velocity component. The method uses focusing along the flow direction to produce signals that are influenced by the shift of the scatterer's position. The signals are then cross-correllated to find the shift in position and thereby the velocity. The performance of the method is investigated using both a flow phantom and in-vivo measurements.
A flow phantom capable of producing a parabolic flow profile was measured with a B-K Medical 8804 7.5 MHz linear array transducer. A plastic tube with an entrance length of 130 cm and a diameter of 17 mm. was used with an EcoWatt 1 pump generating a volume flow of 93.4 l/h corresponding to a peak velocity in the tube of 0.23 m/s. The volume flow was determined by a Danfoss MAG 1100 flow meter. The velocity profiles were measured for different beam-to-flow angles of 90, 65, and 45 degrees. A Harming apodized beam focused at the vessel was transmitted using 64 elements and the received signals on all elements were sampled at 40 MHz and 12 bits in parallel using our experimental ultrasound scanner. Three hundred and seventy pulse echo measurements were acquired for each angle at a pulse repetition frequency of 5 kHz. The field in a number of points on lines parallel to the flow was calculated by focusing the 64 channels;of data. A mean parabolic velocity profile was obtained for purely transverse flow with a mean bias to the true profile of -2.5% relative to the peak velocity and a standard deviation of 13.3% relative to the peak velocity. Twenty pulse-echo lines were used for each estimate and 18 profiles were obtained. For a beam-to-flow angles of 45 and 65 degrees the corresponding numbers were a mean relative bias of less than 4.0% and a relative standard deviation of 3.0% or less, when using 10 pulse-echo lines and 36 profiles. In-vivo measurements have also been performed on the carotid artery on a male volunteer and the flow at an angle of 701 was successfully estimated.
A flow phantom capable of producing a parabolic flow profile was measured with a B-K Medical 8804 7.5 MHz linear array transducer. A plastic tube with an entrance length of 130 cm and a diameter of 17 mm. was used with an EcoWatt 1 pump generating a volume flow of 93.4 l/h corresponding to a peak velocity in the tube of 0.23 m/s. The volume flow was determined by a Danfoss MAG 1100 flow meter. The velocity profiles were measured for different beam-to-flow angles of 90, 65, and 45 degrees. A Harming apodized beam focused at the vessel was transmitted using 64 elements and the received signals on all elements were sampled at 40 MHz and 12 bits in parallel using our experimental ultrasound scanner. Three hundred and seventy pulse echo measurements were acquired for each angle at a pulse repetition frequency of 5 kHz. The field in a number of points on lines parallel to the flow was calculated by focusing the 64 channels;of data. A mean parabolic velocity profile was obtained for purely transverse flow with a mean bias to the true profile of -2.5% relative to the peak velocity and a standard deviation of 13.3% relative to the peak velocity. Twenty pulse-echo lines were used for each estimate and 18 profiles were obtained. For a beam-to-flow angles of 45 and 65 degrees the corresponding numbers were a mean relative bias of less than 4.0% and a relative standard deviation of 3.0% or less, when using 10 pulse-echo lines and 36 profiles. In-vivo measurements have also been performed on the carotid artery on a male volunteer and the flow at an angle of 701 was successfully estimated.
| Original language | English |
|---|---|
| Title of host publication | 2001 IEEE Ultrasonics Symposium Proceedings |
| Volume | 2 |
| Publisher | IEEE |
| Publication date | 2001 |
| Pages | 1405-1408 |
| ISBN (Print) | 0-7803-7177-1 |
| DOIs | |
| Publication status | Published - 2001 |
| Event | 2001 IEEE Ultrasonics Symposium - Atlanta, United States Duration: 7 Oct 2001 → 10 Oct 2001 http://ieeexplore.ieee.org/xpl/mostRecentIssue.jsp?punumber=7781 |
Conference
| Conference | 2001 IEEE Ultrasonics Symposium |
|---|---|
| Country/Territory | United States |
| City | Atlanta |
| Period | 07/10/2001 → 10/10/2001 |
| Internet address |
| Series | I E E E International Ultrasonics Symposium. Proceedings |
|---|---|
| ISSN | 1051-0117 |
Bibliographical note
Copyright: 2000 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEEFingerprint
Dive into the research topics of 'Experimental investigation of transverse velocity estimation using cross-correlation'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver