Estimation of pulses in ultrasound B-scan images

    Research output: Contribution to journalJournal articleResearchpeer-review

    531 Downloads (Orbit)

    Abstract

    It is shown, based on an expression for the received pressure field in pulsed medical ultrasound systems, that a common one-dimensional pulse can be estimated from individual A-lines. An autoregressive moving average (ARMA) model is suggested for the pulse, and an estimator based on the prediction error method is derived. The estimator is used on a segment of an A-line, assuming that the pulse does not change significantly inside the segment. Several examples of the use of the estimator on synthetic data measured from a tissue phantom and in vitro data measured from a calf's liver are given. They show that a pulse can be estimated even at moderate signal-to-noise ratios
    Original languageEnglish
    JournalI E E E Transactions on Medical Imaging
    Volume10
    Issue number2
    Pages (from-to)164-172
    ISSN0278-0062
    DOIs
    Publication statusPublished - 1991

    Bibliographical note

    Copyright: 1991 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEE