Estimation of pulses in ultrasound B-scan images

Research output: Contribution to journalJournal articleResearchpeer-review

338 Downloads (Pure)

Abstract

It is shown, based on an expression for the received pressure field in pulsed medical ultrasound systems, that a common one-dimensional pulse can be estimated from individual A-lines. An autoregressive moving average (ARMA) model is suggested for the pulse, and an estimator based on the prediction error method is derived. The estimator is used on a segment of an A-line, assuming that the pulse does not change significantly inside the segment. Several examples of the use of the estimator on synthetic data measured from a tissue phantom and in vitro data measured from a calf's liver are given. They show that a pulse can be estimated even at moderate signal-to-noise ratios
Original languageEnglish
JournalI E E E Transactions on Medical Imaging
Volume10
Issue number2
Pages (from-to)164-172
ISSN0278-0062
DOIs
Publication statusPublished - 1991

Bibliographical note

Copyright: 1991 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEE