Abstract
The main results in Poulsen & Oakley (2009) are given as the equivalent threshold sound pressure level, ETSPL, measured in an acoustic coupler specifi ed in IEC 60318-3. These results are all correct. The ETSPL values for the ear simulator specifi ed in IEC 60318-1 were calculated from acoustic laboratory measurements and unfortunately a simple sign error has been found that changes two of the columns in Table 3 of Poulsen & Oakley (2009).
| Original language | English |
|---|---|
| Journal | International Journal of Audiology |
| Volume | 53 |
| Issue number | 4 |
| Pages (from-to) | 287-288 |
| ISSN | 1499-2027 |
| DOIs |
|
| Publication status | Published - 2014 |
Keywords
- Audiometry
- Calibration
- ETSPL
- HDA 280
- Hearing threshold
Fingerprint
Dive into the research topics of 'Erratum: Correction to Table 3, in: Equivalent threshold sound pressure levels (ETSPL) for Sennheiser HDA 280 supra-aural audiometric earphones in the frequency range 125 Hz to 8000 Hz (International Journal of Audiology (2009) 48 (271-276))'. Together they form a unique fingerprint.Cite this