Dosimetry intercomparisons in European medical device sterilization plants

A. Miller, P.H.G. Sharpe

    Research output: Contribution to journalJournal articleResearchpeer-review

    Abstract

    Dosimetry intercomparisons have been carried out involving two-thirds of all European radiation sterilization facilities. Dosimeters for the intercomparisons were supplied by two accredited calibration laboratories. The results show good agreement, and indicate overall dosimetry accuracy of the order of +/-5% (1 sigma) for both Co-60 and electron beam plants. (C) 2000 Elsevier Science Ltd. All rights reserved.
    Original languageEnglish
    JournalRadiation Physics and Chemistry
    Volume59
    Issue number3
    Pages (from-to)323-327
    ISSN0969-806X
    DOIs
    Publication statusPublished - 2000