TY - JOUR
T1 - Development and validation of a recombinant human TNF-α based ELISA to detect and quantify adalimumab
AU - Saini, Dinesh Kumar
AU - Devaramani, Manjunath S.
AU - Shanmugavel, Hemalakshmi
AU - Tabassum, Syeda Zuhin
AU - Mudnakudu-Nagaraju, Kiran Kumar
AU - Siddesha, Jalahalli Mariswamy
AU - Shetty, Radhakrishna
PY - 2025
Y1 - 2025
N2 - Adalimumab, a humanized IgG1 monoclonal antibody is currently used to treat inflammatory diseases. However, a sensitive, in-house ELISA for evaluating inter- and intra-individual pharmacokinetic variability of adalimumab remains limited. In this study, an ELISA was developed to measure adalimumab levels, using recombinant human TNF-α (rhTNF-α) as capture antibody. Initially, surface plasma resonance showed acceptable binding kinetics (KD) of 2.38x10−07 nM for adalimumab. Next, a standard curve of adalimumab (1.54 ng/ml to 300 ng/ml), with five quality control points (5.2, 16, 27, 150, and 200 ng/ml) was evaluated for inter and intra-assay accuracy and precision, using serum matrix, by four independent validations. The linear range of the validated assay was 5.2 ng/ml to 200 ng/ml, upper limit of quantification (ULOQ) and lower limit of quantification (LLOQ) were 200 ng/ml and 5.2 ng/ml, respectively. The assay specificity was validated by testing cross-reactivity of rituximab with rhTNF-α, which was found to be non-reactive. Further, the hook effect was over-ruled by diluting the highest concentration of adalimumab tested to assay linear range, and dilution integrity was observed for entire concentrations within linear range (%RE ≤ 20 %), as recommended by European Medicines Agency. Collectively, this rhTNF-α binding-based ELISA method is highly sensitive, reproducible, and useful for monitoring adalimumab.
AB - Adalimumab, a humanized IgG1 monoclonal antibody is currently used to treat inflammatory diseases. However, a sensitive, in-house ELISA for evaluating inter- and intra-individual pharmacokinetic variability of adalimumab remains limited. In this study, an ELISA was developed to measure adalimumab levels, using recombinant human TNF-α (rhTNF-α) as capture antibody. Initially, surface plasma resonance showed acceptable binding kinetics (KD) of 2.38x10−07 nM for adalimumab. Next, a standard curve of adalimumab (1.54 ng/ml to 300 ng/ml), with five quality control points (5.2, 16, 27, 150, and 200 ng/ml) was evaluated for inter and intra-assay accuracy and precision, using serum matrix, by four independent validations. The linear range of the validated assay was 5.2 ng/ml to 200 ng/ml, upper limit of quantification (ULOQ) and lower limit of quantification (LLOQ) were 200 ng/ml and 5.2 ng/ml, respectively. The assay specificity was validated by testing cross-reactivity of rituximab with rhTNF-α, which was found to be non-reactive. Further, the hook effect was over-ruled by diluting the highest concentration of adalimumab tested to assay linear range, and dilution integrity was observed for entire concentrations within linear range (%RE ≤ 20 %), as recommended by European Medicines Agency. Collectively, this rhTNF-α binding-based ELISA method is highly sensitive, reproducible, and useful for monitoring adalimumab.
KW - Adalimumab
KW - Arthritis
KW - Inflammation
KW - ELISA
KW - Pharmacokinetics
KW - Bioanalysis
KW - TNF-Alpha
U2 - 10.1016/j.bbrep.2025.102005
DO - 10.1016/j.bbrep.2025.102005
M3 - Journal article
SN - 2405-5808
SN - 2405-5808
VL - 42
JO - Biochemistry and Biophysics Reports
JF - Biochemistry and Biophysics Reports
M1 - 102005
ER -