Determination of formaldehyde in frozen fish with formaldehyde dehydrogenase using a flow injection system with an incorporated gel-filtration chromatrography column

Iben Ell Bechmann

Research output: Contribution to journalJournal articleResearchpeer-review

Abstract

A flow injection analysis (FIA) system for determination of formaldehyde in frozen fish products is described. The system provides a rapid and selective determination of formaldehyde in aqueous fish extracts by the combination of a deproteinization procedure and a stopped-now enzymatic approach in a FIA system. The FIA system is furnished with a gel-filtration chromatography column for on-line removal of the proteins from the extract before the enzymatic analysis is performed. Compared with the standard methods for determination of formaldehyde in fish products the present method is much faster and less affected by interferences. The limit of detection for the proposed method is 2.5 mg/l of formaldehyde. The sampling frequency is about 10 determinations per hour.
Original languageEnglish
JournalAnalytica Chimica Acta
Volume320
Issue number2-3
Pages (from-to)155-164
ISSN0030-2670
DOIs
Publication statusPublished - 1996

Fingerprint

Dive into the research topics of 'Determination of formaldehyde in frozen fish with formaldehyde dehydrogenase using a flow injection system with an incorporated gel-filtration chromatrography column'. Together they form a unique fingerprint.