Design, calibration and application of broad-range optical nanosensors for determining intracellular pH.

Research output: Contribution to journalJournal articleResearchpeer-review

Abstract

Particle-based nanosensors offer a tool for determining the pH in the endosomal-lysosomal system of living cells. Measurements providing absolute values of pH have so far been restricted by the limited sensitivity range of nanosensors, calibration challenges and the complexity of image analysis. This protocol describes the design and application of a polyacrylamide-based nanosensor (∼60 nm) that covalently incorporates two pH-sensitive fluorophores, fluorescein (FS) and Oregon Green (OG), to broaden the sensitivity range of the sensor (pH 3.1-7.0), and uses the pH-insensitive fluorophore rhodamine as a reference fluorophore. The nanosensors are spontaneously taken up via endocytosis and directed to the lysosomes where dynamic changes in pH can be measured with live-cell confocal microscopy. The most important focus areas of the protocol are the choice of pH-sensitive fluorophores, the design of calibration buffers, the determination of the effective range and especially the description of how to critically evaluate results. The entire procedure typically takes 2-3 weeks.
Original languageEnglish
JournalNature Protocols
Volume9
Issue number12
Pages (from-to)2841-2858
ISSN1745-2473
DOIs
Publication statusPublished - 2014

Fingerprint

Dive into the research topics of 'Design, calibration and application of broad-range optical nanosensors for determining intracellular pH.'. Together they form a unique fingerprint.

Cite this