D-mu-A new concept in industrial low-energy electron dosimetry

Jakob Helt-Hansen, Arne Miller, Peter Sharpe, Bengt Laurell, Doug Weiss, Gary Pageau

    Research output: Contribution to journalJournal articleResearchpeer-review

    Abstract

    Irradiation with low-energy electrons (100–300 keV) results in dose gradients across the thickness of the dosimeters that are typically used for dose measurement at these energies. This leads to different doses being measured with different thickness dosimeters irradiated at the same electron beam, resulting in difficulties in providing traceable dose measurements using reference dosimeters. In order to overcome these problems a new concept is introduced of correcting all measured doses to the average dose in the first micrometer—Dμ. We have applied this concept to dose measurements with dosimeters of different thickness at two electron accelerators operating over a range of energies. The uncertainties of the dose measurements were evaluated, and it was shown that the dose in terms of Dμ was the same at each energy for all dosimeters within the measurement uncertainty. Using the concept of Dμ it is therefore possible to calibrate and measure doses from low-energy electron irradiations with measurement traceability to national standards.
    Original languageEnglish
    JournalRadiation Physics and Chemistry
    Volume79
    Issue number1
    Pages (from-to)66-74
    ISSN0969-806X
    DOIs
    Publication statusPublished - 2010

    Keywords

    • Radiation research and nuclear technologies
    • Radiation physics