Curves for the response of a Ge(Li) detector to gamma rays in the energy range up to 11 MeV

J. Kopecký, W. Ratyński, Inge Elisabeth Warming

    Research output: Contribution to journalJournal articleResearchpeer-review

    Abstract

    The response function of a Ge(Li) coaxial detector with a sensitive volume of 17 cm3 for gamma rays of energies ranging from 2.23 to 10.83 MeV has been determined. The measurements were carried out with an experimental set-up using the neutron beam from the DR 3 reactor ar Risö and the (n, γ) reaction as gamma-ray source. The important features of the spectra are discussed, and the double-escape peak efficiency is determined.
    Original languageEnglish
    JournalNuclear Instruments and Methods
    Volume50
    Issue number2
    Pages (from-to)333-339
    ISSN0029-554X
    DOIs
    Publication statusPublished - 1967