Corrosion of a Fassier-Duval Telescopic Nail Causing Pain and Osteolysis: A Case Report

David Ngo, Maya Todd, Franck Accadbled, Bruce Foster, Morten Stendahl Jellesen, Jan Duedal Rölfing, Jaideep Rawat

Research output: Contribution to journalJournal articleResearchpeer-review

Abstract

CASE: A 15-year-old girl known with osteogenesis imperfecta presented with left femoral pain. She had been treated with multiple Fassier-Duval intramedullary nails, which were still in situ. Radiographic imaging demonstrated focal osteolysis and periosteal reaction at the telescopic junction of the rod in the distal femur. She underwent implant removal. Intraoperative sampling demonstrating acute sterile inflammation and presence of brownish colored particles consistent with metal debris and osteolysis. Explant analysis confirmed corrosion of the stainless-steel telescopic nail as the underlying cause. CONCLUSION: Osteolysis and periosteal reaction because of corrosion should be considered in conjunction with other more common causes of pain, such as fracture or infection, in patients treated with telescopic intramedullary nails.
Original languageEnglish
Article numbere23.00424
JournalJBJS Case Connector
Volume14
Issue number1
ISSN2160-3251
DOIs
Publication statusPublished - 2024

Fingerprint

Dive into the research topics of 'Corrosion of a Fassier-Duval Telescopic Nail Causing Pain and Osteolysis: A Case Report'. Together they form a unique fingerprint.

Cite this