TY - JOUR
T1 - Calibrating Double-Loop H-Field Probe Measurements of RF Coil Current for MRI
AU - Wang, Wenjun
AU - Sánchez-Heredia, Juan Diego
AU - Maurouard, Tom
AU - Zhurbenko, Vitaliy
AU - Ardenkjær-Larsen, Jan Henrik
PY - 2023
Y1 - 2023
N2 - Objective: Double-loop H-field probes are often used to measure current on loop antennae for magnetic resonance imaging (MRI). Loop crosstalk limits the dynamic range of direct measurements with such probes. The crosstalk can be removed by simple calibration. This work analyses the quantitative relation of a probe's calibrated S21 with the RF coil current. Method: The analytical relation between RF coil current and calibrated S21 measurements of a probe is established with the multi-port network theory, and verified by full-wave simulation and benchtop measurements. The effect of calibration is demonstrated by measuring the 1H trap frequency, the active detuning, and the preamplifier decoupling. Results: The calibration removes the effect of crosstalk in a probe and improves the lower bound of |S21|. The calibrated S21 is proportional to coil current. In the lower frequency range, the ratio of calibrated S21 to coil current changes almost linearly with frequency. Impact: The calibration method improves the sensitivity of probe measurements and facilitates fine-tuning current-suppressing circuits like active detuning circuits, traps, preamplifier decoupling. The linear frequency dependency between S21 measurements and coil current allows easy, fair comparison of coil current up to 128 MHz, and in some cases 298 MHz, helping build multi-nucleus coils.
AB - Objective: Double-loop H-field probes are often used to measure current on loop antennae for magnetic resonance imaging (MRI). Loop crosstalk limits the dynamic range of direct measurements with such probes. The crosstalk can be removed by simple calibration. This work analyses the quantitative relation of a probe's calibrated S21 with the RF coil current. Method: The analytical relation between RF coil current and calibrated S21 measurements of a probe is established with the multi-port network theory, and verified by full-wave simulation and benchtop measurements. The effect of calibration is demonstrated by measuring the 1H trap frequency, the active detuning, and the preamplifier decoupling. Results: The calibration removes the effect of crosstalk in a probe and improves the lower bound of |S21|. The calibrated S21 is proportional to coil current. In the lower frequency range, the ratio of calibrated S21 to coil current changes almost linearly with frequency. Impact: The calibration method improves the sensitivity of probe measurements and facilitates fine-tuning current-suppressing circuits like active detuning circuits, traps, preamplifier decoupling. The linear frequency dependency between S21 measurements and coil current allows easy, fair comparison of coil current up to 128 MHz, and in some cases 298 MHz, helping build multi-nucleus coils.
KW - Probes
KW - Coils
KW - Calibration
KW - Current measurement
KW - Magnetic resonance imaging
U2 - 10.1109/JERM.2023.3274742
DO - 10.1109/JERM.2023.3274742
M3 - Journal article
SN - 2469-7257
SP - 1
EP - 7
JO - IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology
JF - IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology
ER -