S. Basso*, B. Salmaso, D. Spiga, M. Ghigo, G. Vecchi, G. Sironi, V. Cotroneo, P. Conconi, E. Redaelli, A. Bianco, G. Pareschi, G. Tagliaferri, D. Sisana, C. Pelliciari, M. Fiorini, S. Incorvaia, M. Uslenghi, L. Paoletti, C. Ferrari, R. LolliA. Zappettini, M. Sanchez del Rio, G. Parodi, V. Burwitz, S. Rukdee, G. Hartner, T. Müller, T. Schmidt, A. Langmeier, D. Della Monica Ferreira, S. Massahi, N. C. Gellert, F. Christensen, M. Bavdaz, I. Ferreira

*Corresponding author for this work

Research output: Chapter in Book/Report/Conference proceedingArticle in proceedingsResearchpeer-review

20 Downloads (Pure)

Abstract

BEaTriX (Beam Expander Testing X-ray facility) is the new facility available at the INAF-Osservatorio Astronomico Brera (Merate, Italy) for the calibration of X-ray optics. Specifically designed to measure the point spread function (PSF) and the effective area (EA) of the mirror modules (MM) of the ATHENA X-ray telescope at their production rate, BEaTriX gives the unique possibility to test the optics with a source that approximate an astronomical source, i.e. with a large, parallel X-ray beam (170 × 60 mm2) that fully covers the entrance pupil of the MM. For a fast and precise testing, BEaTriX is a compact facility (9 × 18 m2) with fast vacuum pump-down (to 10-6 mbar), and an optical setup able to create the X-ray beam with a residual divergence of about 2-3 arcsec, HEW, and with a flux of 60 photons/s/cm2. The first beam line at the energy of 4.51 keV is now operative, and a second beam line, working at the energy of 1.49 keV, will be implemented in the coming future. The unique characteristics of the BEaTriX X-ray beam are obtained with an X-ray microfocus source placed in the focus of a paraboloidal mirror, a monochromation stage with 4 symmetrically cut crystals, and an expansion stage where the beam is diffracted and expanded by an asymmetrically-cut crystal. The beam, reflected by the MMs, is then imaged at 12 m distance, where a directly-illuminated CCD camera is placed. This paper presents the facility, the calibration of the beam and the latest results with the ATHENA MMs.

Original languageEnglish
Title of host publicationProceedings of SPIE : International Conference on Space Optics - ICSO 2022
EditorsKyriaki Minoglou, Nikos Karafolas, Bruno Cugny
PublisherSPIE
Publication date2023
Article number1277720
Chapter10
ISBN (Electronic)9781510668034
DOIs
Publication statusPublished - 2023
EventInternational Conference on Space Optics - Dubrovnik, Croatia
Duration: 3 Oct 20227 Oct 2022

Conference

ConferenceInternational Conference on Space Optics
Country/TerritoryCroatia
CityDubrovnik
Period03/10/202207/10/2022
SeriesProceedings of SPIE - The International Society for Optical Engineering
Volume12777
ISSN0277-786X

Keywords

  • Asymmetric diffraction
  • ATHENA
  • Beam expander
  • BEaTriX
  • Crystals
  • X-ray microfocus source
  • X-ray testing

Fingerprint

Dive into the research topics of 'BEaTriX, the new facility to measure the modular X-ray optics of the ATHENA telescope with an expanded and parallel X-ray beam'. Together they form a unique fingerprint.

Cite this