Automatic Detection of B-lines in In-Vivo Lung Ultrasound

Research output: Contribution to journalJournal article – Annual report year: 2019Researchpeer-review

Documents

DOI

View graph of relations

This paper proposes an automatic method for accurate detection and visualization of B-lines in ultrasound lung scans, which provides a quantitative measure for the number of Blines present. All the scans used in this study were acquired using a BK3000 ultrasound scanner (BK Ultrasound, Denmark) driving a 5.5 MHz linear transducer (BK Ultrasound). Four healthy subjects and four patients after major surgery with pulmonary edema, were scanned at four locations on each lung for B-line examination. Eight sequences of 50 frames were acquired for each subject yielding 64 sequences in total. The proposed algorithm was applied to all 3200 in-vivo lung ultrasound images. The results showed that the average number of B-lines was 0.28±0.06 (Mean±Std) in scans belonging to the patients compared to 0.03±0.06 (Mean±Std) in the healthy subjects. Also, the Mann- Whitney test showed a significant difference between the two groups with p-value of 0.015, and indicating that the proposed algorithm was able to differentiate between the healthy volunteers and the patients. In conclusion, the method can be used to automatically and to quantitatively characterize the distribution of B-lines for diagnosing pulmonary edema.
Original languageEnglish
JournalIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
Volume66
Issue number2
Pages (from-to)309-317
ISSN0885-3010
DOIs
Publication statusPublished - 2019
CitationsWeb of Science® Times Cited: No match on DOI

    Research areas

  • B-lines, Segmentation, Pulmonary edema, Ultrasound imaging

Download statistics

No data available

ID: 162529041