Analysis of Precision of Activation Analysis Method

Kaj Heydorn, K. Nørgaard

    Research output: Contribution to journalJournal articleResearchpeer-review

    Abstract

    The precision of an activation-analysis method prescribes the estimation of the precision of a single analytical result. The adequacy of these estimates to account for the observed variation between duplicate results from the analysis of different samples and materials, is tested by the statistic T, which is shown to be approximated by a χ2 distribution. Application of this test to the results of determinations of manganese in human serum by a method of established precision, led to the detection of airborne pollution of the serum during the sampling process. The subsequent improvement in sampling conditions was shown to give not only increased precision, but also improved accuracy of the results.
    Original languageEnglish
    JournalTalanta
    Volume20
    Issue number9
    Pages (from-to)835-842
    ISSN0039-9140
    DOIs
    Publication statusPublished - 1973