Abstract
An uncomplicated, rapid, automated procedure for the analysis of low cyanide concentrations in whole blood is reported. The analysis was performed by headspace gas chromatography and mass spectrometry in the (1H12C14N) and m/z 29 (1H13C15N). Carryover from cyanide adsorption onto the surface of the needle was prevented by developing a new method that enabled automated flushing of the needle in between each cyanide analysis. Results were compared of ordinary calibrations and those of isotope dilutions. The total time of analysis was 18min for a single cyanide analysis.
| Original language | English |
|---|---|
| Journal | Analytical Letters |
| Volume | 41 |
| Issue number | 14 |
| Pages (from-to) | 2564-2586 |
| ISSN | 0003-2719 |
| DOIs | |
| Publication status | Published - 2008 |
Keywords
- Blood
- Gas chromatography
- Headspace
- Mass spectrometry