A Robust Cryogenic RF Coil (88K) for In-vivo Hyperpolarized 13C MRI of Rats

Juan Diego Sanchez, Rafael Antonio Baron, Esben S. Szocska Hansen, Daniel Højrup Johansen, Vitaliy Zhurbenko, Christoffer Laustsen, Jan Henrik Ardenkjær-Larsen

Research output: Contribution to conferenceConference abstract for conferenceResearchpeer-review

70 Downloads (Pure)

Abstract

We report the performance of a cryogenic RF receive-only coil for 13C imaging of small animals. It is experimentally demonstrated 2-fold SNR improvement in comparison to a room temperature coil in immediate vicinity of the sample. The self-developed cryostat employed for coil cooling shows thermal stability within 5 h of use for 6 L of LN2 , which can be extended up to 12 h if more LN is added. The Q88K /Q290K ratio of
the unloaded coil is 550/285
Original languageEnglish
Publication date2019
Number of pages2
Publication statusPublished - 2019
EventISMRM 27th Annual Meeting & Exhibition - Palais des congrès de Montréal, Montréal, Canada
Duration: 11 May 201916 May 2019
http://www.ismrm.org/19m/

Conference

ConferenceISMRM 27th Annual Meeting & Exhibition
LocationPalais des congrès de Montréal
CountryCanada
CityMontréal
Period11/05/201916/05/2019
Internet address

Cite this

Sanchez, J. D., Baron, R. A., Hansen, E. S. S., Johansen, D. H., Zhurbenko, V., Laustsen, C., & Ardenkjær-Larsen, J. H. (2019). A Robust Cryogenic RF Coil (88K) for In-vivo Hyperpolarized 13C MRI of Rats. Abstract from ISMRM 27th Annual Meeting & Exhibition, Montréal, Canada.