Abstract
We report the performance of a cryogenic RF receive-only coil for 13C imaging of small animals. It is experimentally demonstrated 2-fold SNR improvement in comparison to a room temperature coil in immediate vicinity of the sample. The self-developed cryostat employed for coil cooling shows thermal stability within 5 h of use for 6 L of LN2 , which can be extended up to 12 h if more LN is added. The Q88K /Q290K ratio of
the unloaded coil is 550/285
the unloaded coil is 550/285
Original language | English |
---|---|
Publication date | 2019 |
Number of pages | 2 |
Publication status | Published - 2019 |
Event | ISMRM 27th Annual Meeting & Exhibition - Palais des congrès de Montréal, Montréal, Canada Duration: 11 May 2019 → 16 May 2019 http://www.ismrm.org/19m/ |
Conference
Conference | ISMRM 27th Annual Meeting & Exhibition |
---|---|
Location | Palais des congrès de Montréal |
Country/Territory | Canada |
City | Montréal |
Period | 11/05/2019 → 16/05/2019 |
Internet address |