A new luminescence detection and stimulation head for the Risø TL/OSL reader

    Research output: Contribution to journalConference articleResearchpeer-review

    427 Downloads (Pure)

    Abstract

    A new automated Detection And Stimulation Head (DASH) has been developed for the Risø TL/OSL luminescence reader to provide easy access to new technologies, new signals and new measurement methods. The automated DASH includes a filter changer and a detector changer that makes it possible to change stimulation filters (4×4 filter combinations possible) and detectors (3 detectors possible) as part of a measurements sequence. The new automated DASH with dedicated driver electronics does not affect the use of other attachments, and can be retrospectively fitted to existing Risø TL/OSL readers.
    Original languageEnglish
    JournalRadiation Measurements
    Volume81
    Pages (from-to)178–184
    ISSN1350-4487
    DOIs
    Publication statusPublished - 2015
    Event14th International Conference on Luminescence and Electron Spin Resonance Dating - Montréal, Canada
    Duration: 7 Jul 201511 Jul 2015
    Conference number: 14

    Conference

    Conference14th International Conference on Luminescence and Electron Spin Resonance Dating
    Number14
    CountryCanada
    CityMontréal
    Period07/07/201511/07/2015

    Keywords

    • Instrumentation
    • Optical stimulation
    • Spatially resolved detection
    • Automation
    • Technology transfer
    • Automated detection
    • Driver electronics
    • Luminescence detection
    • Luminescence reader
    • New Measurement Method
    • Spatially resolved
    • Luminescence

    Fingerprint Dive into the research topics of 'A new luminescence detection and stimulation head for the Risø TL/OSL reader'. Together they form a unique fingerprint.

    Cite this