3-D Directional Transverse Oscillations Synthetic Aperture Vector Flow Imaging with a 1024 Element Matrix Probe

Research output: Chapter in Book/Report/Conference proceedingArticle in proceedings – Annual report year: 2019Researchpeer-review

Documents

View graph of relations

3-D vector flow imaging is simulated with a 1024 element matrix array and vessels aligned with the x- and yaxes. Parabolic flow profiles with a peak velocity of 0.5 m/s are
simulated with an interleaved synthetic aperture sequence with 5 emissions at a pulse repetition frequency of 15 kHz. Two-peak receive apodizations are used to induce transverse oscillations (TO) in the x- and y-directions, and a directional TO velocity estimator based on cross-correlation is used to estimate the 3 velocity components. The relative mean bias and relative mean standard deviation are calculated for each velocity component and the velocity magnitude. For the vessel aligned with the x-axis,
the bias of the x-component is -10.1%, and for the vessel aligned with the y-axis, the bias of the y-component is -2.04%. Relative mean standard deviations of the x- and y-components are in the range 10% to 17% and below 3.2% for the z-component.
Original languageEnglish
Title of host publicationProceedings of 2019 IEEE International Ultrasonics Symposium
Number of pages4
PublisherIEEE
Publication statusAccepted/In press - 2019
Event2019 IEEE International Ultrasonics Symposium - SEC Glasgow, Glasgow, United Kingdom
Duration: 6 Oct 20199 Oct 2019
http://attend.ieee.org/ius-2019/

Conference

Conference2019 IEEE International Ultrasonics Symposium
LocationSEC Glasgow
CountryUnited Kingdom
CityGlasgow
Period06/10/201909/10/2019