High-speed counting and sizing of cells in an impedance flow microcytometer with compact electronic instrumentation

Research output: Contribution to journalJournal article – Annual report year: 2014Researchpeer-review

View graph of relations

Here we describe a high-throughput impedance flow cytometer on a chip. This device was built using compact and inexpensive electronic instrumentation. The system was used to count and size a mixed cell sample containing red blood cells and white blood cells. It demonstrated a counting capacity of up to ~500 counts/s and was validated through a synchronised high-speed optical detection system. In addition, the device showed excellent discrimination performance under high-throughput conditions.
Original languageEnglish
JournalMicrofluidics and Nanofluidics
Volume16
Issue number1-2
Pages (from-to)91-99
ISSN1613-4982
DOIs
Publication statusPublished - 2014
CitationsWeb of Science® Times Cited: No match on DOI