Volume dose ratios relevant for alanine dosimetry in small, 6 MV photon beams
Publication: Research - peer-review › Journal article – Annual report year: 2012
Standard
Volume dose ratios relevant for alanine dosimetry in small, 6 MV photon beams. / Cronholm, Rickard O.; Andersen, Claus Erik; Behrens, Claus F.; Helt-Hansen, Jakob.
In: Radiation Measurements, Vol. 47, No. 10, 2012, p. 1014-1017.Publication: Research - peer-review › Journal article – Annual report year: 2012
Harvard
APA
CBE
MLA
Vancouver
Author
Bibtex
}
RIS
TY - JOUR
T1 - Volume dose ratios relevant for alanine dosimetry in small, 6 MV photon beams
A1 - Cronholm,Rickard O.
A1 - Andersen,Claus Erik
A1 - Behrens,Claus F.
A1 - Helt-Hansen,Jakob
AU - Cronholm,Rickard O.
AU - Andersen,Claus Erik
AU - Behrens,Claus F.
AU - Helt-Hansen,Jakob
PB - Pergamon
PY - 2012
Y1 - 2012
N2 - The overall objective of this work is to establish alanine dosimetry for traceable measurements in clinical radiotherapy beams, in particular for non-reference situations such as small field sizes and composite beam delivery (e.g. intensity modulated radiotherapy and volumetric modulated arc therapy). To this end, we here present the results of a Monte Carlo simulation study with DOSRZnrc that investigated the influence of field and detector size for small 6 MV photon beams. The study focusses on doses averaged over the volume of the detector rather than point doses.The ratio of volume averaged doses to water (D¯W) and alanine (D¯det) was found to be approximately 1.025 for most situations studied, and a constant ratio is likely to be representative for many applications in radiation therapy. However, D¯W/D¯det was found to be as low as 0.9908 ± 0.0037 in situations where one might expect significant deviations from charged particle equilibrium (i.e. at shallow depths and when the field size was smaller than the range of the secondary electrons). These effects therefore need consideration when finite-size alanine dosimeters are used under such conditions.
AB - The overall objective of this work is to establish alanine dosimetry for traceable measurements in clinical radiotherapy beams, in particular for non-reference situations such as small field sizes and composite beam delivery (e.g. intensity modulated radiotherapy and volumetric modulated arc therapy). To this end, we here present the results of a Monte Carlo simulation study with DOSRZnrc that investigated the influence of field and detector size for small 6 MV photon beams. The study focusses on doses averaged over the volume of the detector rather than point doses.The ratio of volume averaged doses to water (D¯W) and alanine (D¯det) was found to be approximately 1.025 for most situations studied, and a constant ratio is likely to be representative for many applications in radiation therapy. However, D¯W/D¯det was found to be as low as 0.9908 ± 0.0037 in situations where one might expect significant deviations from charged particle equilibrium (i.e. at shallow depths and when the field size was smaller than the range of the secondary electrons). These effects therefore need consideration when finite-size alanine dosimeters are used under such conditions.
KW - Alanine
KW - Small field
KW - Correction factor
KW - Monte Carlo
U2 - 10.1016/j.radmeas.2012.07.009
DO - 10.1016/j.radmeas.2012.07.009
JO - Radiation Measurements
JF - Radiation Measurements
SN - 1350-4487
IS - 10
VL - 47
SP - 1014
EP - 1017
ER -