Improved Internal Reference Oxygen Sensors with Composite Ceramic Electrodes

Publication: Research - peer-reviewJournal article – Annual report year: 2012

Documents

DOI

View graph of relations

Potentiometric oxygen sensors with an internal reference electrode, which uses the equilibrium pO2 of the binary mixture of Ni/NiO as the reference, are demonstrated. The cells employ Pt or composite ceramics as the sensing electrode. The cells are fabricated by a flexible and potentially low cost method and show high accuracy. The errors in cell voltage are less than 4 mV above 359°C. The cells exhibit good stability (>5100 hrs), extended working temperature range, good tolerance to thermal and pO2 cycling, fast response and easy recoverability when Ni is depleted. Both cell fabrication and performance are highly reproducible. The composite ceramics, based on strontium doped manganite and yttria doped zirconia, are proven superior over Pt to serve as the electrode material.
Original languageEnglish
JournalElectrochemical Society. Journal
Publication date2012
Volume159
Issue10
PagesB811-B817
ISSN0013-4651
DOIs
StatePublished
CitationsWeb of Science® Times Cited: 3
Download as:
Download as PDF
Select render style:
APAAuthorCBEHarvardMLAStandardVancouverShortLong
PDF
Download as HTML
Select render style:
APAAuthorCBEHarvardMLAStandardVancouverShortLong
HTML
Download as Word
Select render style:
APAAuthorCBEHarvardMLAStandardVancouverShortLong
Word

Download statistics

No data available

ID: 10590231