Standard

Equivalent threshold sound pressure levels (ETSPL) for Sennheiser HDA 280 supra-aural audiometric earphones in the frequency range 125 Hz to 8000 Hz. / Poulsen, Torben; Oakley, Sebastian.

In: International Journal of Audiology, Vol. 48, No. 5, 2009, p. 271-276.

Publication: Research - peer-reviewJournal article – Annual report year: 2009

Harvard

APA

CBE

MLA

Vancouver

Author

Poulsen, Torben; Oakley, Sebastian / Equivalent threshold sound pressure levels (ETSPL) for Sennheiser HDA 280 supra-aural audiometric earphones in the frequency range 125 Hz to 8000 Hz.

In: International Journal of Audiology, Vol. 48, No. 5, 2009, p. 271-276.

Publication: Research - peer-reviewJournal article – Annual report year: 2009

Bibtex

@article{97ef5504c88c47a3bb75b90b50663932,
title = "Equivalent threshold sound pressure levels (ETSPL) for Sennheiser HDA 280 supra-aural audiometric earphones in the frequency range 125 Hz to 8000 Hz",
keywords = "HDA 280, Audiometry, ETSPL, Hearing threshold, Calibration",
publisher = "Informa Healthcare",
author = "Torben Poulsen and Sebastian Oakley",
year = "2009",
doi = "10.1080/14992020902788982",
volume = "48",
number = "5",
pages = "271--276",
journal = "International Journal of Audiology",
issn = "1499-2027",

}

RIS

TY - JOUR

T1 - Equivalent threshold sound pressure levels (ETSPL) for Sennheiser HDA 280 supra-aural audiometric earphones in the frequency range 125 Hz to 8000 Hz

A1 - Poulsen,Torben

A1 - Oakley,Sebastian

AU - Poulsen,Torben

AU - Oakley,Sebastian

PB - Informa Healthcare

PY - 2009

Y1 - 2009

N2 - Hearing threshold sound pressure levels were measured for the Sennheiser HDA 280 audiometric earphone. Hearing thresholds were measured for 25 normal hearing test subjects at the 11 audiometric test frequencies from 125 Hz to 8000 Hz. Sennheiser HDA 280 is a supra-aural earphone that may be seen as a substitute for the classical Telephonics TDH 39. The results are given as the Equivalent Threshold Sound Pressure Level, ETSPL, measured in an acoustic coupler specified in IEC 60318-3. The results are in good agreement with an independent investigation from PTB, Braunschweig, Germany. From acoustic laboratory measurements, ETSPL values are calculated for the ear simulator specified in IEC 60318-1. Fitting of earphone and coupler is discussed. The data may be used for a future update of the RETSPL standard for supra-aural audiometric earphones ISO 389-1.

AB - Hearing threshold sound pressure levels were measured for the Sennheiser HDA 280 audiometric earphone. Hearing thresholds were measured for 25 normal hearing test subjects at the 11 audiometric test frequencies from 125 Hz to 8000 Hz. Sennheiser HDA 280 is a supra-aural earphone that may be seen as a substitute for the classical Telephonics TDH 39. The results are given as the Equivalent Threshold Sound Pressure Level, ETSPL, measured in an acoustic coupler specified in IEC 60318-3. The results are in good agreement with an independent investigation from PTB, Braunschweig, Germany. From acoustic laboratory measurements, ETSPL values are calculated for the ear simulator specified in IEC 60318-1. Fitting of earphone and coupler is discussed. The data may be used for a future update of the RETSPL standard for supra-aural audiometric earphones ISO 389-1.

KW - HDA 280

KW - Audiometry

KW - ETSPL

KW - Hearing threshold

KW - Calibration

U2 - 10.1080/14992020902788982

DO - 10.1080/14992020902788982

JO - International Journal of Audiology

JF - International Journal of Audiology

SN - 1499-2027

IS - 5

VL - 48

SP - 271

EP - 276

ER -